Abstract
This paper describes the on-line sample pretreatment and analysis of proteins and peptides with a poly(methylmethacrylate) (PMMA) microfluidic device (IonChip). This chip consists of two hyphenated electrophoresis channels with integrated conductivity detectors. The first channel can be used for sample preconcentration and sample clean-up, while in the second channel the selected compounds are separated. Isotachophoresis (ITP) combined with zone electrophoresis (CZE) was used to preconcentrate a myoglobin sample by a factor of about 65 before injection into the second dimension and to desalt a mixture of six proteins with 100 mM NaCl. However, ITP-CZE could not be used for the removal of two proteins from a protein/peptide sample since the protein zone in the ITP step was too small to remove certain compounds. Therefore, we used CZE-CZE for the removal of proteins from a protein/peptide mixture, thereby injecting only the peptides into the second CZE separation channel.
| Original language | Undefined/Unknown |
|---|---|
| Pages (from-to) | 68-73 |
| Number of pages | 6 |
| Journal | Journal of Chromatography A |
| Volume | 839 |
| Issue number | 1-2 |
| Publication status | Published - 2006 |
UN SDGs
This output contributes to the following UN Sustainable Development Goals (SDGs)
-
SDG 3 Good Health and Well-being
Keywords
- Farmacie/Biofarmaceutische wetenschappen (FARM)
- Medical technology
- Farmacie(FARM)
- Bescherming en bevordering van de menselijke gezondheid
- Biomedische technologie en medicijnen
- Public Health
- Analytical chemistry
- Drugmisbruik en verslaving