Abstract
OBJECTIVE: To assess intra- and interobserver repeatability of ocular biometric measurements obtained by means of high-resolution B-mode ultrasonography in dogs. Animals-6 Beagles without ocular abnormalities.
PROCEDURES: B-mode ultrasonography was performed bilaterally with a 10.5-MHz broadband compact linear array transducer. All measurements were made on 2 different occasions by 2 observers. The Bland-Altman method was used to assess agreement between measurements obtained by the 2 observers and between the 2 sets of measurements obtained by each observer.
RESULTS: Intra- and interobserver repeatability was highest for larger measurements, such as depth of the eye and depth of the anterior chamber. When repeatability was examined, bias was significantly different from 0 for only a few measurements, but the percentage difference between observations was as high as 180% for some measurements.
CONCLUSIONS AND CLINICAL RELEVANCE: Results suggest that most measurements of intraocular distances and structures obtained by means of high-resolution B-mode ultrasonography have acceptable intra- and interobserver repeatability. However, the percentage difference between observations can be high for smaller measurements.
| Original language | English |
|---|---|
| Pages (from-to) | 1743-9 |
| Number of pages | 7 |
| Journal | American Journal of Veterinary Research |
| Volume | 67 |
| Issue number | 10 |
| DOIs | |
| Publication status | Published - 2006 |
Keywords
- Animals
- Biometry
- Dogs
- Eye
- Female
- Male
- Observer Variation
- Reproducibility of Results
Fingerprint
Dive into the research topics of 'Intraobserver and interobserver repeatability of ocular biometric measurements obtained by means of B-mode ultrasonography in dogs'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver