T1 - Flow variability and its physical causes in infusion technology
T2 - a systematic review of in vitro measurement and modeling studies
AU - Snijder, Roland A
AU - Konings, Maurits K
AU - Lucas, Peter
AU - Egberts, Toine C
AU - Timmerman, Annemoon D
PY - 2015/8
Y1 - 2015/8
N2 - Infusion therapy is medically and technically challenging and frequently associated with medical errors. When administering pharmaceuticals by means of infusion, dosing errors can occur due to flow rate variability. These dosing errors may lead to adverse effects. We aimed to systematically review the available biomedical literature for in vitro measurement and modeling studies that investigated the physical causes of flow rate variability. Special focus was given to syringe pump setups, which are typically used if very accurate drug delivery is required. We aimed to extract from literature the component with the highest mechanical compliance in syringe pump setups. We included 53 studies, six of which were theoretical models, two articles were earlier reviews of infusion literature, and 45 were in vitro measurement studies. Mechanical compliance, flow resistance, and dead volume of infusion systems were stated as the most important and frequently identified physical causes of flow rate variability. The syringe was indicated as the most important source of mechanical compliance in syringe pump setups (9.0×10-9 to 2.1×10-8 l/Pa). Mechanical compliance caused longer flow rate start-up times (from several minutes up to approximately 70 min) and delayed occlusion alarm times (up to 117 min).
AB - Infusion therapy is medically and technically challenging and frequently associated with medical errors. When administering pharmaceuticals by means of infusion, dosing errors can occur due to flow rate variability. These dosing errors may lead to adverse effects. We aimed to systematically review the available biomedical literature for in vitro measurement and modeling studies that investigated the physical causes of flow rate variability. Special focus was given to syringe pump setups, which are typically used if very accurate drug delivery is required. We aimed to extract from literature the component with the highest mechanical compliance in syringe pump setups. We included 53 studies, six of which were theoretical models, two articles were earlier reviews of infusion literature, and 45 were in vitro measurement studies. Mechanical compliance, flow resistance, and dead volume of infusion systems were stated as the most important and frequently identified physical causes of flow rate variability. The syringe was indicated as the most important source of mechanical compliance in syringe pump setups (9.0×10-9 to 2.1×10-8 l/Pa). Mechanical compliance caused longer flow rate start-up times (from several minutes up to approximately 70 min) and delayed occlusion alarm times (up to 117 min).
U2 - 10.1515/bmt-2014-0148
DO - 10.1515/bmt-2014-0148
M3 - Article
C2 - 26352348
SN - 0013-5585
VL - 60
SP - 277
EP - 300
JO - Biomedizinische Technik
JF - Biomedizinische Technik
IS - 4
ER -