Skip to main navigation Skip to search Skip to main content

Development and validation of an LC–MS/MS method for the quantification of the KRASG12C inhibitor divarasib

  • Jamie Rijmers
  • , Viët Bui
  • , Maria C. Lebre
  • , Alfred H. Schinkel
  • , Jos H. Beijnen
  • , Olaf van Tellingen
  • , L. J. van Winden*
  • *Corresponding author for this work
  • Netherlands Cancer Institute

Research output: Contribution to journal › Article › Academic › peer-review

Abstract

Divarasib is a newly developed covalent KRASG12C inhibitor, currently under clinical investigation in a phase 3 trial in patients with non-small cell lung cancer (NSCLC). At the moment, very limited pharmacokinetic data are publicly known. However, obtaining more insight into the pharmacokinetic properties of divarasib is important, since this may provide a better understanding of its efficacy and safety risks. Pre-clinical studies have been performed in mouse models to evaluate the effect of drug transporters and drug-metabolizing enzymes on the plasma exposure and tissue distribution of divarasib. Therefore, a reliable quantification method is required. To our knowledge, no bioanalytical assay of divarasib has been published yet. Therefore, in this study we developed and validated an assay to quantify divarasib in human plasma and in eight different mouse-related matrices, and partially in mouse plasma, using liquid chromatography-tandem mass spectrometry (LC–MS/MS). The method was initially evaluated over a concentration range of 1–10,000 nM. However, due to carry-over observed at 10,000 nM, the validated calibration range was established at 1–2000 nM, with matrix-dependent LLOQs of 1–10 nM. Erlotinib was used as an internal standard and acetonitrile was utilized to perform protein precipitation as sample pretreatment. Divarasib demonstrated stability in human plasma and in mouse plasma and tissue homogenates under various experimental conditions. A pilot in vivo study showed the applicability of our validated LC–MS/MS method. Ongoing clinical trials may collect plasma samples, and this developed method enables quantification of divarasib in both mouse and human plasma samples.

Original languageEnglish
Article number125211
JournalJournal of Chromatography B: Analytical Technologies in the Biomedical and Life Sciences
Volume1282
DOIs
Publication statusPublished - 1 Oct 2026

Bibliographical note

Publisher Copyright:
© 2026 Published by Elsevier B.V.

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well-being

Keywords

  • Bioanalysis
  • Divarasib
  • KRAS
  • LC–MS/MS
  • Mouse matrices

Fingerprint

Dive into the research topics of 'Development and validation of an LC–MS/MS method for the quantification of the KRASG12C inhibitor divarasib'. Together they form a unique fingerprint.

Cite this