Determination of platinum originating from carboplatin in human urine and canine excretion products by inductively coupled plasma mass spectrometry.

T. Janssens, E.E.M. Brouwers, J.P. de Vos, J.H.M. Schellens, J.H. Beijnen

Research output: Contribution to journalArticleAcademicpeer-review

Abstract

We present highly sensitive, rapid methods for the determination of Pt originating from carboplatin in human urine and canine urine, feces, and oral fluid. The methods are based on the quantification of Pt by inductively coupled plasma mass spectrometry, and allow quantification of 7.50 ng/L Pt in human and canine urine (in 15 muL of matrix), 15.0 ng/L Pt in canine oral fluid (in 15 muL of matrix), and 0.105 ng/g Pt in canine feces (in 5 mug of matrix). Sample pretreatment mainly involved dilution with appropriate diluents. The performance of the methods fulfilled the most recent FDA guidelines for bioanalytical method validation. Validated ranges of quantification were 7.50 to 1.00 x 10(4) ng/L Pt in human and canine urine, 0.105-30.0 ng/g Pt in canine feces, and 15.0 to 1.00 x 10(4) ng/L Pt in canine oral fluid. Canine urine and oral fluid cannot be easily obtained. Therefore, we also investigated the validity of the usage of human matrix samples for the preparation of calibration standards and quality control samples as alternatives, to be used in future clinical studies. The assays are used to support biomonitoring studies and pharmacokinetic studies in pet dogs treated with carboplatin.
Original languageUndefined/Unknown
Pages (from-to)153-61
Number of pages9
JournalJournal of Analytical Toxicology
Volume35
Issue number3
Publication statusPublished - 2011

Cite this