Abstract
Proton-transfer-reaction mass spectrometry (PTR-MS) is an analytical technique used to monitor volatile organic compounds in real-time. For quantitative analysis, compounds of interest are typically calibrated using gas standards, but PTR-MS is quantitative to uncalibrated compounds if the mass-dependent transmission is well defined. However, long-term measurements are challenging due to the drift in transmission over time. Performing frequent calibrations helps, but the methods are time-consuming and tedious, often leading to instruments being under-calibrated. Here we show the use of long-lived and globally monitored compounds in the atmosphere as a tool to constrain the transmission between calibrations. The major ion of trichlorofluromethane (CFC-11) and carbon tetrachloride (CCl4) is found at the mass-to-charge ratio (m/z) 116.9, which we propose using to retrieve the transmission of a PTR-MS. We determined the pseudo-reaction rate constants of CFC-11 and CCl4 to be 0.82 × 10−9 ± 0.05 × 10−9 cm3 s−1 molecule−1, and 1.65 × 10−9 ± 0.08 × 10−9 cm3 s−1 molecule−1, respectively. The method introduced here can improve data quality and accuracy, especially for long-term atmospheric measurements.
Original language | English |
---|---|
Article number | 117311 |
Pages (from-to) | 1-6 |
Number of pages | 6 |
Journal | International Journal of Mass Spectrometry |
Volume | 504 |
DOIs | |
Publication status | Published - Oct 2024 |
Bibliographical note
Publisher Copyright:© 2024 The Authors
Funding
This work has received funding from the European Union's Horizon 2020 research and innovation programme under the Marie Sklodowska-Curie grant agreement No 955750 and the Ruisdael Observatory, a scientific infrastructure co-financed by the Dutch Research Council (NWO, grant number 184.034.015) .
Funders | Funder number |
---|---|
European Union | 955750 |
Dutch Research Council (NWO) | 184.034.015 |
Keywords
- Calibration
- CCl
- CFC-11
- PTR-MS
- Transmission