Aerosol analysis using a Thermal-Desorption Proton-Transfer-Reaction Mass Spectrometer (TD-PTR-MS): A new approach to study processing of organic aerosols

R. Holzinger, J. Williams, F. Herrmann, J. Lelieveld, N. M. Donahue, T. Roeckmann

Research output: Contribution to journalArticleAcademicpeer-review

Abstract

We present a novel analytical approach to measure the chemical composition of organic aerosol. The new instrument combines proton-transfer-reaction mass-spectrometry (PTR-MS) with a collection-thermal-desorption aerosol sampling technique. For secondary organic aerosol produced from the reaction of ozone with isoprenoids in a laboratory reactor, the TD-PTR-MS instrument detected typically 80% of the mass that was measured with a scanning mobility particle sizer (SMPS). The first field deployment of the instrument was the EUCAARI-IOP campaign at the CESAR tall tower site in the Netherlands. For masses with low background values (∼30% of all masses) the detection limit of aerosol compounds was below 0.2 ng/m3 which corresponds to a sampled compound mass of 35 pg. Comparison of thermograms from ambient samples and from chamber-derived secondary organic aerosol shows that, in general, organic compounds from ambient aerosol samples desorb at much higher temperatures than chamber samples. This suggests that chamber aerosol is not a good surrogate for ambient aerosol and therefore caution is advised when extrapolating results from chamber experiments to ambient conditions
Original languageEnglish
Pages (from-to)2257-2267
Number of pages11
JournalAtmospheric chemistry and physics
Volume10
Issue number5
DOIs
Publication statusPublished - 2010

Fingerprint

Dive into the research topics of 'Aerosol analysis using a Thermal-Desorption Proton-Transfer-Reaction Mass Spectrometer (TD-PTR-MS): A new approach to study processing of organic aerosols'. Together they form a unique fingerprint.

Cite this